Orthopedics — Total Knee Replacement
Freedom<br> Total Knee®
The one and only 7 tangential radii knee implant in the world — an advanced system designed to address the anatomical, physiological and lifestyle needs of patients.
Orthopedics — Total Knee Replacement
The one and only 7 tangential radii knee implant in the world — an advanced system designed to address the anatomical, physiological and lifestyle needs of patients.
Tangential radii
Degrees of high flexion
Approved
01 — The system
The Freedom Total Knee® System is the first USFDA-approved knee system with 7 tangential radii, resulting in excellent implant-to-bone fit.
Developed using advanced design engineering technologies, the system’s unique design allows patients to achieve optimal high-flexion motion.
Compared to general knee replacement procedures, the system requires minimal bone cutting and enables a more conservative bone resection.
Its intuitive design offers surgeons the freedom to customise size options to better fit the implants to the patient’s knee.
02 — Design rationale
The Freedom femoral design is based on functional principles of progressive multi-radii patello-femoral and tibio-femoral contact areas.
01
Radii 1, 2 and 3 ensure optimized patella-femoral contact. Radii 4, 5, 6 and 7 control femoral rollback and flexion over a wide array of biomechanical requirements, from walking to climbing stairs.
02
7 radii along with a thin anterior flange and a 6-degree deep trochlear groove facilitate predictable patello-femoral tracking — helping to reduce anterior knee pain.
03
A multi-radii design gives predictable femoral rollback, allowing high flexion up to 155 degrees for various daily and social activities.
04
A modified post-cam mechanism allows rotation up to 15 degrees at deep flexion, preventing post-cam impingement and reducing wear and tear of the poly.
05
Freedom Knee concentrates load towards the centre of the poly at deep flexion, rather than at the condylar edges.
06
A 5-point locking mechanism and an optimized inside surface roughness averaging 11 µin reduce back-side wear in the tibial tray, improving durability.
07
A thin anterior flange, low-profile box and multi-radii design minimise posterior condylar resection, resulting in maximum bone conservation.
08
5-point cobalt chrome tibial locking system with 3-degree and 30 mm keel; ram-extruded poly UHMWPE, GUR 1020. Both CR and PS tibial inserts share identical locking mechanisms to mate.
09
CR and PS femurs, metal-backed tibial trays and inserts — with compatibility of a 2-size mix and match in critical sizes.
03 — Global shape and sizing
| Femur | A | B | C | D | E | F | G | H |
|---|---|---|---|---|---|---|---|---|
| M/L | 54.00 | 58.00 | 62.00 | 64.00 | 66.00 | 70.00 | 74.00 | 78.00 |
| A/P | 51.00 | 54.00 | 58.00 | 60.00 | 62.00 | 66.00 | 70.00 | 74.00 |
| Tray | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
|---|---|---|---|---|---|---|---|---|
| M/L | 59 | 62 | 66 | 66 | 71 | 72 | 76 | 78 |
| A/P | 40 | 40 | 42 | 46 | 48 | 50 | 52 | 54 |
| Femoral size | A | B | C | D | E | F | G | H |
|---|---|---|---|---|---|---|---|---|
| Post-condylar resection (mm) | 7.2 | 7.7 | 8.2 | 8.5 | 8.8 | 9.4 | 9.8 | 10.4 |
| Box-cut resection (cm³) | 4.37 | 5.42 | 6.44 | 7.14 | 7.91 | 9.69 | 11.26 | 13.44 |
04 — Product details
| Item | Value |
|---|---|
| Product | Freedom Total Knee® System |
| Category | Orthopedics · total knee replacement |
| Femoral geometry | 7 tangential radii · 6° deep trochlear groove |
| Flexion | High flexion up to 155° |
| Components | CR and PS femurs (L & R) · metal-backed tibial trays and inserts · all-poly tibia |
| Materials | Cobalt chrome · ram-extruded UHMWPE, GUR 1020 |
| Patella diameters | 25, 28, 31, 34, 37 and 40 mm |
| Liner thickness | 9, 11, 14 and 17 mm |
| Fixation | Cemented use only |
| Regulatory | USFDA-approved |
Certification & documentation
Full documentation is available to healthcare professionals and professional buyers on request.
| Document | Format |
|---|---|
| Freedom Knee® design rationale | |
| Clinical data | |
| Instructions for Use (IFU) |
Get in touch
Caution: this device is restricted to sale by or on the order of a licensed physician. Warning: this device is intended for cemented use only. Please see the package insert for complete device description, indications, contraindications, precautions, adverse effects, warnings, materials, sterilization and patient guidance associated with the Freedom Total Knee® System.
test